MΩ/cm
3
0 to 19.99 0.01
uS/cm
3
0 to 2.000 0.01
0 to 20.00 0.1
0 to 200.0 0.1
0 to 5000 1.0
mS/cm
3
0 to 20.00 10
0 to 200.0 50
Section I - Specifications Page 2
MULTI-PARAMETER TRANSMITTER USER’S MANUAL
Section I - Specifications
pH ORP Conductivity Flow
Display 2 x 16 alpha-numeric LCD display
Power Requirements 4 to 20mA, Loop Powered, 16 to 32 VDC
Flow: 0 to 9999 with
selectable
flow rate units
pH: 0.01 to 14.00 ORP: -1999 to +1999mV
Volume: 0 - 999 with Auto Range
Measuring Range
(Dependent on sensor)
Flow rate units: Gallons (GP), Cubic
Temp: 0 to 100°C or Temp: 0 to 100°C or
Feet (CF), Liters (LP), Cubic Meters
32° to +212°F 32° to +212°F (
CM), custom by entering factor
related to Gallons
Time units: Seconds (S), Minutes (M)
Temp: 0 to 100°C or 32° to +212°F
Hours (H)
Automatic or Manual
Temperature Automatic or Manual
Not r
equired
User selectable temperature
Not required
Compensation 0 - 100°C (32° to +212°F)
compensation slope 0.0 to 10.0%/°C.
0 to 100°C (32° to +212°F)
Temperature Unit °C or °F Not required
Temperature Sensor User selectable: 300Ω NTC Thermistor, 3000Ω NTC Thermistor or Pt. 1000 RTD Not required
Auto-Calibration
Manual Calibration
Dry Calibration
Calibration Modes Manual Calibration
T
emperature Calibration
Sample Calibration K factor Input
Temperature Calibration Temperature Calibration
Ambient Conditions Temperature: -20°C to +60°C or -4°F to +140°F Humidity: 0 to 90% RH (non-condensing)
Sensor to Transmitter Differential Sensor: 3000 ft
300 ft
2000 ft
Distance Combination Sensor: 10 ft
Analog Output
4 to 20mA
Isolated Output, Range expand 0 to 100% of full scale (min segment 10% of full scale), max. load 800Ω
Memory Back-up All user settings are retained indefinitely in memory (EEPROM)
SHARK
TX Enclosure: NEMA 4X, 1/4 DIN, polycarbonate enclosure with two 1/2” conduit holes
Mechanical
SHARK
TXP Enclosure: NEMA 4X front panel, 1/4 DIN, polycarbonate
SHARK
TX Mounting: Universal Mounting kit for surface, pipe and panel mount included
SHARKTXP Mounting: Panel and DIN rail mount included
Sensor Input
Probe: -600 to +600mV Probe: -1999 to +1999mV Cell: 0 to 9999Ω
P
addle: 0 to 2000Hz
Temp. Sensor: 0 to 9999Ω Temp. Sensor: 0 to 9999Ω Temp. Sensor: 0 to 9999Ω
Invalid Entries Invalid entries cannot be stored
Manual Test Mode Process value can be simulated with arrow keys to verify correct setup of output
Output Hold 4 to 20mA output is placed on hold when the transmitter is in Menu mode
Recall data from last
Recall data from last calibration, calibration mode,
calibration, calibration buffer
Calibration Data
1st & 2nd accepted buffer value and probe mV output,
accepted v
alue, and cell Recall store K factor.
calibration temperature, calibration slope, and probe
resistance, calibration
efficiency
temperature
Auto Return User selectable auto return if the transmitter is left in menu mode for more than 10 min.
Display Damping User can select rate at which the transmitter updates display. Enables display damping of unstable process
Net Weight SHARK
TX: 0.71 lbs (0.32 kg)
SHARKTXP: 0.25 lbs (0.12 kg)
Approvals ULC (pending)